MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-02-07 for CAVITRON PLUS PACKAGE G136 8161425 manufactured by Dentsply Llc.
[179624808]
There has been a previous report received where lack of water flow has caused an overheating insert. Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur. As such, this event meets the criteria for reportability per 21 cfr part 803. Water tubing in hp cable restricted, possible reason for the overheating. No sterimate handpiece was sent in, could be the reason for overheating. Added new cable along with new water filter and foot pedal batteries.
Patient Sequence No: 1, Text Type: N, H10
[179624809]
While using a g136 scaler, there was blocked water flow and the handpiece and water was getting hot; no injury resulted.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2424472-2020-00002 |
MDR Report Key | 9682101 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2020-02-07 |
Date of Report | 2020-02-07 |
Date Mfgr Received | 2020-01-08 |
Date Added to Maude | 2020-02-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | DENTIST |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. KARL NITTINGER |
Manufacturer Street | 221 W. PHILADELPHIA ST. SUITE 60W |
Manufacturer City | YORK PA 17401 |
Manufacturer Country | US |
Manufacturer Postal | 17401 |
Manufacturer Phone | 7178494229 |
Manufacturer G1 | DENTSPLY LLC |
Manufacturer Street | 1301 SMILE WAY |
Manufacturer City | YORK PA 17404 |
Manufacturer Country | US |
Manufacturer Postal Code | 17404 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CAVITRON PLUS PACKAGE |
Generic Name | SCALER, ULTRASONIC |
Product Code | ELC |
Date Received | 2020-02-07 |
Returned To Mfg | 2020-01-15 |
Model Number | G136 |
Catalog Number | 8161425 |
Lot Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | DENTSPLY LLC |
Manufacturer Address | 1301 SMILE WAY YORK PA 17404 US 17404 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-07 |