COCR TRANSCEND? FEMORAL HEAD 28MM SLT TAPER SHORT 26000017

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2020-02-07 for COCR TRANSCEND? FEMORAL HEAD 28MM SLT TAPER SHORT 26000017 manufactured by Microport Orthopedics Inc..

Event Text Entries

[178548212] This event will be updated once the investigation is complete. Trends will be evaluated.
Patient Sequence No: 1, Text Type: N, H10


[178548213] Allegedly, shell did not attach to the ball. Additional information on 01/31/2020: reporter stated that during surgery the head did not attach to the shell. They were using a 28 head and a 45 bipolar shell. According to the reporter, he noticed that the plastic rim inside the shell did not seem flush to the shell (there was a 3-4mm gap) and he thinks that it may be the reason why the head did not attach to the shell. In the end, they used a 32 head and a 46 shell. These functioned well.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010536692-2020-00105
MDR Report Key9682793
Report SourceDISTRIBUTOR
Date Received2020-02-07
Date of Report2020-02-07
Date Facility Aware2020-01-17
Device Manufacturer Date2020-01-17
Date Added to Maude2020-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street5677 AIRLINE ROAD
Manufacturer CityARLINGTON TN 38002
Manufacturer CountryUS
Manufacturer Postal38002
Manufacturer Phone9018674771
Manufacturer G1MICROPORT ORTHOPEDICS INC.
Manufacturer Street5677 AIRLINE RD.
Manufacturer CityARLINGTON TN 38002
Manufacturer CountryUS
Manufacturer Postal Code38002
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOCR TRANSCEND? FEMORAL HEAD 28MM SLT TAPER SHORT
Generic NameHIP COMPONENT
Product CodeJDL
Date Received2020-02-07
Returned To Mfg2020-01-17
Model Number26000017
Catalog Number26000017
Lot Number1760412
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROPORT ORTHOPEDICS INC.
Manufacturer Address5677 AIRLINE RD. ARLINGTON TN 38002 US 38002


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-07