MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-07 for ENSITE VELOCITY? SYSTEM VELOCITY AMPLIFIER 100014514 manufactured by St. Jude Medical, Inc..
[178202305]
The results, method and conclusion codes along with investigation results will be provided in the final report.
Patient Sequence No: 1, Text Type: N, H10
[178202306]
During a procedure, the system displayed an amplifier error. After validation was performed, the system lost the connection with the amplifier. The issue was initially resolved by power cycling the amplifier, but once ablation was started, the error displayed again. Troubleshooting included changing the cables and restarting the amplifier, but the issue remained. There were no adverse consequences to the patient but the procedure was cancelled.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2184149-2020-00018 |
MDR Report Key | 9683087 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-02-07 |
Date of Report | 2020-03-24 |
Date of Event | 2020-02-06 |
Date Mfgr Received | 2020-03-19 |
Device Manufacturer Date | 2010-07-08 |
Date Added to Maude | 2020-02-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | STEPHANIE O' SULLIVAN |
Manufacturer Street | 5050 NATHAN LANE NORTH |
Manufacturer City | PLYMOUTH MN 55442 |
Manufacturer Country | US |
Manufacturer Postal | 55442 |
Manufacturer Phone | 6517565400 |
Manufacturer G1 | ST. JUDE MEDICAL, INC. |
Manufacturer Street | ONE ST. JUDE MEDICAL DRIVE |
Manufacturer City | ST. PAUL MN 55117 |
Manufacturer Country | US |
Manufacturer Postal Code | 55117 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ENSITE VELOCITY? SYSTEM VELOCITY AMPLIFIER |
Generic Name | COMPUTER, DIAGNOSTIC, PROGRAMMABLE |
Product Code | DQK |
Date Received | 2020-02-07 |
Returned To Mfg | 2020-03-18 |
Model Number | 100014514 |
Catalog Number | 100014514 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ST. JUDE MEDICAL, INC. |
Manufacturer Address | ONE ST. JUDE MEDICAL DRIVE ST. PAUL MN 55117 US 55117 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-07 |