MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-07 for SELENIA MAMMOGRAPHY SYSTEM SEL-00014 manufactured by Hologic, Inc.
[182562339]
It was reported that the c-arm has begun to drift downwards without noise or errors. No injury reported. A field engineer was dispatched to the site and was not able to reproduce the reported issue of drifting downwards. The customer also reported that there was noise using the c-arm rotation and the field engineered cleaned the stereoloc brake plate to resolve this issue. Once this was completed the system was working as intended.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1220984-2020-00017 |
| MDR Report Key | 9683759 |
| Report Source | USER FACILITY |
| Date Received | 2020-02-07 |
| Date of Report | 2020-01-17 |
| Date of Event | 2020-01-17 |
| Date Mfgr Received | 2020-01-17 |
| Date Added to Maude | 2020-02-07 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | KRISTIN FORNIERI |
| Manufacturer Street | 36 & 37 APPLE RIDGE ROAD |
| Manufacturer City | DANBURY CT 06810 |
| Manufacturer Country | US |
| Manufacturer Postal | 06810 |
| Manufacturer Phone | 2037318491 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | SELENIA MAMMOGRAPHY SYSTEM |
| Generic Name | FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM |
| Product Code | MUE |
| Date Received | 2020-02-07 |
| Model Number | SEL-00014 |
| Catalog Number | SEL-00014 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | N |
| Device Eval'ed by Mfgr | Y |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | HOLOGIC, INC |
| Manufacturer Address | 36 & 37 APPLE RIDGE ROAD DANBURY CT 06810 US 06810 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2020-02-07 |