PHILIPS M1026B ANESTHETIC GAS MODULE FGPH1026B

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2020-02-07 for PHILIPS M1026B ANESTHETIC GAS MODULE FGPH1026B manufactured by Philips Medizin Systeme Boeblingen Gmbh.

Event Text Entries

[178965787] The complaint device was returned for evaluation. Device sent in due to the device shutting down in the middle of a procedure. Incoming visual inspection found no anomalies. Device evaluation found no problem. The device was calibrated, the case and front/rear connectors were inspected. The circuit boards were inspected. The reported event was unable to be duplicated and there was no problem found; therefore, a root cause cannot be determined. This type of event will continue to be monitored. It should be noted that this is being filed based on retrospective review.
Patient Sequence No: 1, Text Type: N, H10


[178965788] Reportedly, post repair, the device failed during a procedure as it shut off. There was no report of patient harm or intervention. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007409280-2020-00040
MDR Report Key9684663
Report SourceCONSUMER,HEALTH PROFESSIONAL
Date Received2020-02-07
Date of Report2018-07-23
Date Mfgr Received2018-07-23
Date Added to Maude2020-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJENNA VARGAS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal92672
Manufacturer Phone8004495328
Manufacturer G1AVANTE HEALTH SOLUTIONS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal Code92672
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePHILIPS M1026B ANESTHETIC GAS MODULE
Generic NameANESTHETIC GAS MODULE
Product CodeCBQ
Date Received2020-02-07
Returned To Mfg2018-08-01
Model NumberFGPH1026B
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDIZIN SYSTEME BOEBLINGEN GMBH
Manufacturer AddressHEWLETT-PACKARD-STR.-2 BOEBLINGEN, 71034 GM 71034


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-07

© 2026 FDA.report