CONSULT HCG URINE CASSETTE 5001 25T FHC-A102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2020-02-07 for CONSULT HCG URINE CASSETTE 5001 25T FHC-A102 manufactured by Alere San Diego.

Event Text Entries

[188841483] Results pending completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


[188841484] The distributer reported that a women's health facility had false negative results on the hcg urine cassette initially providing a lot number. For one patient, the result for four tests with one specimen was 2 false positive results and 2 negative results with a negative serum test result. Subsequent contact with the customer revealed that the customer could not actually identify which lot number was used when the inaccurate results were obtained. The customer was not willing to provide additional details but would send the product back. On 01/22/2020, lot numbers hcg9072026, hcg9040123, hcg9040084 were received. It is still unknown which lot was used at the time of the events. For this reason, one mdr will report this event. This mdr is associated with 2027969-2020-0009.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2020-00010
MDR Report Key9684778
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2020-02-07
Date of Report2020-03-31
Date Mfgr Received2020-03-30
Date Added to Maude2020-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMONIKA BURRELL
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052506
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCONSULT HCG URINE CASSETTE 5001 25T
Generic NameHCG PREGNANCY TEST
Product CodeJHI
Date Received2020-02-07
Returned To Mfg2020-03-23
Model NumberFHC-A102
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO
Manufacturer Address9995 SUMMERS RIDGE RD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-07

© 2026 FDA.report