[186832178] Pending evaluation. Concomitant device(s): humeral adapter tray (cat#: 320-10-00 / sn#: (b)(4). 42mm humeral liner (cat#: 320-42-00 / sn#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186832179] As reported, the initial implant of the (b)(6) male patient's left shoulder was (b)(6) 2019. The patient complained of "clicking" sound in his left shoulder to surgeon some time after surgery. After x-raying the shoulder the surgeon discovered that the humeral platform tray had separated from the humeral stem with humeral liner still attached to the tray. The patient was then scheduled for a left shoulder revision on (b)(6) 2020. The surgeon revised the patient's shoulder by replacing the humeral tray, torque screw, and humeral liner. Upon initial inspection of failed implants there did not appear to be any obvious signs of failure. Possible worn down/stripped threads on the end of the torque screw. Replaced torque screw and humeral tray had no issues implanting and fixating into the humeral stem. Patient was last known to be in stable condition following the event. Product will be returned for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1038671-2020-00036
MDR Report Key9687443
Report SourceHEALTH PROFESSIONAL
Date Received2020-02-10
Date of Report2020-02-10
Date of Event2020-02-03
Date Mfgr Received2020-02-03
Device Manufacturer Date2019-08-05
Date Added to Maude2020-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATE JACOBSON
Manufacturer Phone3523771140
Manufacturer G1EXACTECH, INC.
Manufacturer Street2320 NW 66 CT
Manufacturer CityGAINESVILLE FL 32653
Manufacturer CountryUS
Manufacturer Postal Code32653
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEQUINOXE
Generic NameREVERSE TORQUE DEFINING SCREW KIT
Product CodeKWT
Date Received2020-02-10
Model NumberUNK
Catalog Number320-20-00
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerEXACTECH, INC.
Manufacturer Address2320 NW 66 COURT GAINESVILLE FL 32653 US 32653


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-02-10