NASOGASTRIC SUMP TUBE WITH PREVENT? ANTI-REFLUX FILTER AND LOPEZ VALVE? 0056160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,other,user facili report with the FDA on 2020-02-10 for NASOGASTRIC SUMP TUBE WITH PREVENT? ANTI-REFLUX FILTER AND LOPEZ VALVE? 0056160 manufactured by C.r. Bard, Inc. (covington) -1018233.

Event Text Entries

[182799731] The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
Patient Sequence No: 1, Text Type: N, H10


[182799732] It was reported that the diameter of the connection on the ng tube is small which makes decompression of the stomach more difficult. The customer noted that the patient was in the hospital for small bowel obstruction and had a ng tube in for decompression. The patient was very nauseated for several hours, unrelieved by medication and the ng tube was bringing back up some gastric content but not a lot. The provider reportedly hand irrigated the ng tube very forcefully with at least 2 syringes of water and after the second syringe, the ng tube dumped out 750ml of gastric contents.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1018233-2020-00924
MDR Report Key9688408
Report SourceDISTRIBUTOR,OTHER,USER FACILI
Date Received2020-02-10
Date of Report2020-03-06
Date Mfgr Received2020-03-03
Date Added to Maude2020-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYONIC ANDERSON
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal30014
Manufacturer Phone7707846100
Manufacturer G1C.R. BARD, INC. (COVINGTON) -1018233
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal Code30014
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNASOGASTRIC SUMP TUBE WITH PREVENT? ANTI-REFLUX FILTER AND LOPEZ VALVE?
Generic NameNG TUBE
Product CodeFEG
Date Received2020-02-10
Model Number0056160
Catalog Number0056160
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerC.R. BARD, INC. (COVINGTON) -1018233
Manufacturer Address8195 INDUSTRIAL BLVD COVINGTON GA 30014 US 30014


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.