ACRYSOF IQ TORIC SINGLEPIECE IOL SN6AT4 SN6AT4.085

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-10 for ACRYSOF IQ TORIC SINGLEPIECE IOL SN6AT4 SN6AT4.085 manufactured by Alcon Laboratories Ireland Ltd..

Event Text Entries

[178540848] Evaluation summary: the product was not returned for analysis. The product history and batch records were reviewed and documentation indicated the product met release criteria. The product investigation could not identify a root cause. There have been no other complaints reported in the lot number. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[178540849] A health professional reported that during an intraocular lens (iol) implant procedure, the haptic broke and did not come out of the cartridge during implantation. Additional information received stating the lens went into the eye and the surgeon had to widen the incision to remove the lens. A suture was used to close the wound. The surgeon did not have another lens to implant so the procedure was not completed. The surgeon completed the procedure the next day with no further complications.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612169-2020-00050
MDR Report Key9688763
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-10
Date of Report2020-02-10
Date of Event2020-01-22
Date Mfgr Received2020-02-07
Device Manufacturer Date2018-05-28
Date Added to Maude2020-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JASON MICHAELIDES
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175686438
Manufacturer G1ALCON LABORATORIES IRELAND LTD.
Manufacturer StreetCORK BUSINESS&TECHNOLOGY PARK MODEL FARM ROAD
Manufacturer CityCORK 00000
Manufacturer CountryEI
Manufacturer Postal Code00000
Single Use3
Remedial ActionNO
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACRYSOF IQ TORIC SINGLEPIECE IOL
Generic NameLENS, INTRAOCULAR, TORIC OPTICS
Product CodeMJP
Date Received2020-02-10
Model NumberSN6AT4
Catalog NumberSN6AT4.085
Lot Number21216563
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON LABORATORIES IRELAND LTD.
Manufacturer AddressCORK BUSINESS&TECHNOLOGY PARK MODEL FARM ROAD CORK 00000 EI 00000


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-10

© 2026 FDA.report