BD VACUTAINER? Z (NO ADDITIVE) PLUS URINE TUBE 364915

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,other report with the FDA on 2020-02-10 for BD VACUTAINER? Z (NO ADDITIVE) PLUS URINE TUBE 364915 manufactured by Becton, Dickinson And Company (bd).

Event Text Entries

[187076988] (b)(6). A device evaluation is anticipated, but has not yet begun. Upon completion of the investigation, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[187076989] It was reported that a cracked tube occurred before use with a bd vacutainer? Z (no additive) plus urine tube. The following information was provided by the initial reporter, "it was reported during bd internal testing that the tubes failed drop test under ratings 3 and 4 which are considered to have exposure to blood borne pathogens. Bd vacutainer? Urine tubes catalog # 364915. As part of capa (b)(4), we found that above mentioned catalogs had failures of drop test under ratings 3 & 4 which are considered to have exposure to blood borne pathogens. Each catalog had 300 nominal dosed samples and 300 maximum dosed samples for drop testing. All the samples are tested. There is no root cause found for these failures. Irradiation dose (kgy): maximum (35. 6-42. 6)" 5 occurrences were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9617032-2020-00104
MDR Report Key9689130
Report SourceCOMPANY REPRESENTATIVE,OTHER
Date Received2020-02-10
Date of Report2020-02-19
Date of Event2019-05-29
Date Mfgr Received2020-01-20
Device Manufacturer Date2019-04-15
Date Added to Maude2020-02-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BECTON, DICKINSON AND COMPANY (BD)
Manufacturer StreetBELLIVER WAY BELLIVER INDUSTRIAL ESTATE
Manufacturer CityPLYMOUTH
Manufacturer CountryUK
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBD VACUTAINER? Z (NO ADDITIVE) PLUS URINE TUBE
Generic NameTRANSPORT CULTURE MEDIUM
Product CodeJSM
Date Received2020-02-10
Returned To Mfg2020-01-22
Catalog Number364915
Lot Number9105862
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECTON, DICKINSON AND COMPANY (BD)
Manufacturer AddressBELLIVER WAY BELLIVER INDUSTRIAL ESTATE PLYMOUTH UK


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-02-10

© 2024 FDA.report