ACRYSOF IQ TORIC SINGLEPIECE IOL SN6AT4 SN6AT4.205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-11 for ACRYSOF IQ TORIC SINGLEPIECE IOL SN6AT4 SN6AT4.205 manufactured by Alcon Laboratories Ireland Ltd..

Event Text Entries

[182564773] The product was returned for analysis and the reported damage was observed. Iol returned in two segments in the iol case. Solution is dried on both segments. One haptic is broken/torn and is returned. The optic is torn/split-cut dividing the iol in two and scratched/marked-rejectable. We are unable to determine the root cause for the reported complaint "defective". The returned iol shows evidence of possible handling by the customer due to the presence of solution dried on both segments. In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria. Based on these investigation findings, we are unable to verify if the iol contributed to the event. All product and batch history records are quality reviewed prior to product release. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[182564774] A health professional reported that two intraocular lens (iol) implants were defective. Additional information was requested but not received.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612169-2020-00051
MDR Report Key9693413
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-11
Date of Report2020-02-11
Date Mfgr Received2020-02-11
Device Manufacturer Date2017-09-18
Date Added to Maude2020-02-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SCHLECH
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514979
Manufacturer G1ALCON LABORATORIES IRELAND LTD.
Manufacturer StreetCORK BUSINESS&TECHNOLOGY PARK MODEL FARM ROAD
Manufacturer CityCORK 00000
Manufacturer CountryEI
Manufacturer Postal Code00000
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACRYSOF IQ TORIC SINGLEPIECE IOL
Generic NameLENS, INTRAOCULAR, TORIC OPTICS
Product CodeMJP
Date Received2020-02-11
Returned To Mfg2020-01-02
Model NumberSN6AT4
Catalog NumberSN6AT4.205
Lot Number21201091
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALCON LABORATORIES IRELAND LTD.
Manufacturer AddressCORK BUSINESS&TECHNOLOGY PARK MODEL FARM ROAD CORK 00000 EI 00000