THERMOPHORE 155

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-02-11 for THERMOPHORE 155 manufactured by Battle Creek Equipment.

Event Text Entries

[182722336] Customer stated, " the pad caught fire by the wires leading into the switch. " customer did not claim injury. Product was not returned. An investigation into the manufactured lot found no other instances of potential fires occurring from this lot. An investigation into returned products for pad fire found that fire was primarily caused by customer misuse of twisting/ wrapping the cord to tight or folding/ laying on the pad. The ifu states, "loop cord loosely when storing. Tight wrapping may damage cord and internal parts. ", "do not sit on, lie on, or crush pad. Avoid sharp folds". Further investigation will be conducted once the product is returned, along with a supplemental report submitted. Updated: product was returned 1/28/2020 and investigated on 1/29/2020. The investigator observed evidence of fire/ burn cut on the cord. Additionally the investigator observed that the cord was twisted. The investigator determined that the twisting of the cord lead to a breaking of the casing by the switch, which then caused the customers reported event. The ifu states, "loop cord loosely when storing. Tight wrapping may damage cord and internal parts. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1832415-2020-11020
MDR Report Key9696204
Report SourceCONSUMER
Date Received2020-02-11
Date Added to Maude2020-02-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL GREEN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHERMOPHORE
Generic NameTHERMOPHORE
Product CodeIRT
Date Received2020-02-11
Returned To Mfg2020-01-28
Model Number155
Lot Number00155065195
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBATTLE CREEK EQUIPMENT
Manufacturer Address702 S REED RD FREMONT, IN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-11

© 2026 FDA.report