3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) 3100 A 768901

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-12 for 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) 3100 A 768901 manufactured by Vyaire Medical.

Event Text Entries

[183135267] The field service representative performed and completed several calibrations to this unit such as airway pressure monitor transducer, pneumatic check and/or adjustment, alarm function test, driver displacement indicator adjustment, driver controller adjustment, pulse width modulator (pwm) adjustment, patient circuit calibration and a performance test. There was no problem detected as the unit operates within specification. The field service representative recommends that the device must undergo for 12k/7yr preventive maintenance. Any additional information received from the customer will be included in a follow-up report.
Patient Sequence No: 1, Text Type: N, H10


[183135268] The customer reported all the settings on 3100a ventilator were drifting including the frequency while ventilating on a patient. The patient was transferred to another ventilator. At this time, patient harm is unknown.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2020-11386
MDR Report Key9701923
Report SourceUSER FACILITY
Date Received2020-02-12
Date of Report2020-01-15
Date of Event2020-01-15
Date Mfgr Received2020-01-15
Device Manufacturer Date2009-01-22
Date Added to Maude2020-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MINDY FABER
Manufacturer Street26125 N. RIVERWOODS BLVD.
Manufacturer CityMETTAWA IL 60045
Manufacturer CountryUS
Manufacturer Postal60045
Manufacturer Phone8727570116
Manufacturer G1VYAIRE MEDICAL INC.
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Generic NameVENTILATOR, HIGH FREQUENCY
Product CodeLSZ
Date Received2020-02-12
Model Number3100 A
Catalog Number768901
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL
Manufacturer Address26125 N. RIVERWOODS BLVD. METTAWA IL 60045 US 60045


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.