MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-12 for 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) 3100 A 768901 manufactured by Vyaire Medical.
[183135267]
The field service representative performed and completed several calibrations to this unit such as airway pressure monitor transducer, pneumatic check and/or adjustment, alarm function test, driver displacement indicator adjustment, driver controller adjustment, pulse width modulator (pwm) adjustment, patient circuit calibration and a performance test. There was no problem detected as the unit operates within specification. The field service representative recommends that the device must undergo for 12k/7yr preventive maintenance. Any additional information received from the customer will be included in a follow-up report.
Patient Sequence No: 1, Text Type: N, H10
[183135268]
The customer reported all the settings on 3100a ventilator were drifting including the frequency while ventilating on a patient. The patient was transferred to another ventilator. At this time, patient harm is unknown.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2021710-2020-11386 |
MDR Report Key | 9701923 |
Report Source | USER FACILITY |
Date Received | 2020-02-12 |
Date of Report | 2020-01-15 |
Date of Event | 2020-01-15 |
Date Mfgr Received | 2020-01-15 |
Device Manufacturer Date | 2009-01-22 |
Date Added to Maude | 2020-02-12 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. MINDY FABER |
Manufacturer Street | 26125 N. RIVERWOODS BLVD. |
Manufacturer City | METTAWA IL 60045 |
Manufacturer Country | US |
Manufacturer Postal | 60045 |
Manufacturer Phone | 8727570116 |
Manufacturer G1 | VYAIRE MEDICAL INC. |
Manufacturer Street | 1100 BIRD CENTER DRIVE |
Manufacturer City | PALM SPRINGS CA 92262 |
Manufacturer Country | US |
Manufacturer Postal Code | 92262 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) |
Generic Name | VENTILATOR, HIGH FREQUENCY |
Product Code | LSZ |
Date Received | 2020-02-12 |
Model Number | 3100 A |
Catalog Number | 768901 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | VYAIRE MEDICAL |
Manufacturer Address | 26125 N. RIVERWOODS BLVD. METTAWA IL 60045 US 60045 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-02-12 |