THERMOPHORE 156

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-02-12 for THERMOPHORE 156 manufactured by Battle Creek Equipment.

Event Text Entries

[182930728] Customer stated, "notice a brown spot on her cover. Removed the cover, and found a burn hole in the heating pad. She can see the coil. Used pad in recliner, sitting up straight with pad behind back. " customer did not claim injury. Product was returned. Investigator observed discoloration of the unit, cover, thermostats, and a discolored hole. Additionally the investigator observed no evidence of burning. The investigator determined that the observed hole was a worn in hole. The investigator determined, along with the customers statement, that the use of the product on the recliner caused bunching of the lower part of the pad, which caused a worn hole and discoloration. Ifu states, "do not sit on, lie on, or crush pad. Avoid sharp folds".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1832415-2020-11094
MDR Report Key9702639
Report SourceCONSUMER
Date Received2020-02-12
Date of Report2020-01-16
Date Added to Maude2020-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL GREEN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHERMOPHORE
Generic NameTHERMOPHORE
Product CodeIRT
Date Received2020-02-12
Returned To Mfg2020-01-21
Model Number156
Lot Number00156115182
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBATTLE CREEK EQUIPMENT
Manufacturer Address702 S REED RD FREMONT, IN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-12

© 2026 FDA.report