[188689818] Evaluation and investigation of the device showed no relevant error which may have caused or contributed to the occurred event (fall as well as change of resistance behavior).
Patient Sequence No: 1, Text Type: N, H10


[188689819] Air in hydraulics, pt. States that walking feels different. Change of resistance behaviour: patient feels insecure. Patient fell and fractured her wrist.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615892-2020-00001
MDR Report Key9706154
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2020-02-13
Date of Report2019-12-12
Date of Event2019-11-23
Date Mfgr Received2020-01-17
Device Manufacturer Date2017-05-10
Date Added to Maude2020-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. REINHARD WOLKERSTORFER
Manufacturer StreetBREHMSTRASSE 16
Manufacturer CityVIENNA, VIENNA 1110
Manufacturer CountryAU
Manufacturer Postal1110
Manufacturer G1OTTO BOCK HEALTHCARE PRODUCTS GMBH
Manufacturer StreetBREHMSTRASSE 16
Manufacturer CityVIENNA, 1110
Manufacturer CountryAU
Manufacturer Postal Code1110
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGENIUM X3
Generic NameGENIUM X3
Product CodeISY
Date Received2020-02-13
Model Number3B5-2
Catalog Number3B5-2
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOTTO BOCK HEALTHCARE PRODUCTS GMBH
Manufacturer AddressBREHMSTRASSE 16 VIENNA, 1110 AU 1110


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-02-13

© 2026 FDA.report