OLYMPUS A22002A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,05,06 report with the FDA on 2007-12-28 for OLYMPUS A22002A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[773125] The user facility reported that the tip of the telescope and a lazer device was noted to be burnt during a transurethral resection of prostate (turp). The facility's staff reported that the attending physician was using a homien laser with the telescope at the time of use. The model, serial number and the manufacturer of the laser device used with the telescope are unknown. The facility's staff indicated that the procedure was completed with the same telescope. There was no allegation of patient harm, and there was no complication reported.
Patient Sequence No: 1, Text Type: D, B5


[7896146] The device referenced in this report was returned to olympus for investigation. The evaluation confirmed evidence of thermal damage to the distal tip of the telescope. Moisture and fluid were noted inside the objective cover glass. There were also broken fibers and sharp edges observed which were determined to be due to physical damage. Additionally, there was deep scrape and bend noted on the outer tube, which was attributed to user handling. This report is being filed as an mdr in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2007-00028
MDR Report Key970626
Report Source00,05,06
Date Received2007-12-28
Date of Report2007-11-30
Date Added to Maude2008-04-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS
Generic NameTELESCOPE
Product CodeFBP
Date Received2007-12-28
Model NumberA22002A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key973585
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressHAMBURG


Patients

Patient NumberTreatmentOutcomeDate
10 2007-12-28

© 2026 FDA.report