MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-13 for INDIGO SYSTEM CATRX ASPIRATION CATHETER CATRXKIT manufactured by Penumbra, Inc..
[180810190]
This device is available for return. A follow up mdr will be submitted upon completion of the device investigation.
Patient Sequence No: 1, Text Type: N, H10
[180810191]
The patient was undergoing a thrombectomy procedure in the right tibial vessels using an indigo system catrx aspiration catheter (catrx), a non-penumbra sheath, and a guidewire. During the procedure, the catrx broke and a segment of the catheter became lodged inside the patient's right superficial femoral artery (sfa). The segment was pulled to the left common femoral artery using a snare and surgical cut down was performed to retrieve the segment. The procedure was then ended. There was no report of an adverse effect to the patient.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 3005168196-2020-00185 |
| MDR Report Key | 9707811 |
| Report Source | COMPANY REPRESENTATIVE,HEALTH |
| Date Received | 2020-02-13 |
| Date of Report | 2020-01-21 |
| Date of Event | 2020-01-21 |
| Report Date | 2005-01-01 |
| Date Reported to FDA | 2005-01-01 |
| Date Reported to Mfgr | 2005-01-10 |
| Date Mfgr Received | 2020-01-21 |
| Date Added to Maude | 2020-02-13 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | MS. VERONICA FARRIS |
| Manufacturer Street | ONE PENUMBRA PLACE |
| Manufacturer City | ALAMEDA CA 94502 |
| Manufacturer Country | US |
| Manufacturer Postal | 94502 |
| Manufacturer Phone | 5107483200 |
| Single Use | 0 |
| Previous Use Code | 0 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | INDIGO SYSTEM CATRX ASPIRATION CATHETER |
| Generic Name | QEX |
| Product Code | QEX |
| Date Received | 2020-02-13 |
| Model Number | CATRXKIT |
| Catalog Number | CATRXKIT |
| Lot Number | F94127 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | PENUMBRA, INC. |
| Manufacturer Address | ONE PENUMBRA PLACE ALAMEDA CA 94502 US 94502 |
| Product Code | --- |
| Date Received | 2020-02-13 |
| Device Sequence No | 101 |
| Device Event Key | 0 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 1. Required No Informationntervention | 2020-02-13 |