CARELINK SMARTSYNC BASE 24970A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-13 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[179415545] Product analysis: analysis of the logs was able to confirm the report that the programmer application closed when triggering the high gain mode under waveform settings. It was confirmed that the malfunction was a result of a known documented issue that was under investigation by the company. It is believed that a fix for the issue was implemented in the next application software version update. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[179415546] It was reported that the programmer application closed and returned to the home screen when attempting to turn on the highgain mode in the analyzer settings. It was noted that connection between the base and patient connector were also lost. After reestablishing connection and staring up the analyzer it showed that highgain mode was on. The application shut down again upon turning the mode off. The issue was escalated for investigation into the root cause. The programmer and application remain in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004593495-2020-00147
MDR Report Key9708133
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-13
Date of Report2020-02-13
Date of Event2020-01-17
Date Mfgr Received2020-01-17
Date Added to Maude2020-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-02-13
Model Number24970A
Catalog Number24970A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-13

© 2024 FDA.report