VIGILEO MHM1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-13 for VIGILEO MHM1 manufactured by Edwards Lifesciences.

Event Text Entries

[188741672] The product is expected to be returned for analysis; however, it has not yet been received. Upon the return of the product a supplemental report will be sent with the investigation results as well as the device history record. Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
Patient Sequence No: 1, Text Type: N, H10


[188741673] As reported, using this vigileo monitor, the displayed stroke volume was 15 or 20 ml/beat. Zeroing was possible but the customer considered the displayed value incorrect and exchanged the monitor. The issue was solved after exchanging the monitor. The expected value was unknown. The patient was not treated based on the incorrect value. There was no patient injury reported. Patient demographic information requested but unavailable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2015691-2020-10506
MDR Report Key9709117
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-13
Date of Report2020-01-21
Date of Event2020-01-21
Date Mfgr Received2020-01-21
Date Added to Maude2020-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SAMANTHA EVELEIGH
Manufacturer StreetONE EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal92614
Manufacturer Phone9492503939
Manufacturer G1EDWARDS LIFESCIENCES LLC
Manufacturer StreetONE EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal Code92614
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIGILEO
Generic NameCOMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION
Product CodeDXG
Date Received2020-02-13
Model NumberMHM1
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerEDWARDS LIFESCIENCES
Manufacturer AddressONE EDWARDS WAY IRVINE CA 92614 US 92614


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-13

© 2026 FDA.report