MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-13 for VIGILEO MHM1 manufactured by Edwards Lifesciences.
[188741672]
The product is expected to be returned for analysis; however, it has not yet been received. Upon the return of the product a supplemental report will be sent with the investigation results as well as the device history record. Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
Patient Sequence No: 1, Text Type: N, H10
[188741673]
As reported, using this vigileo monitor, the displayed stroke volume was 15 or 20 ml/beat. Zeroing was possible but the customer considered the displayed value incorrect and exchanged the monitor. The issue was solved after exchanging the monitor. The expected value was unknown. The patient was not treated based on the incorrect value. There was no patient injury reported. Patient demographic information requested but unavailable.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2015691-2020-10506 |
MDR Report Key | 9709117 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-02-13 |
Date of Report | 2020-01-21 |
Date of Event | 2020-01-21 |
Date Mfgr Received | 2020-01-21 |
Date Added to Maude | 2020-02-13 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. SAMANTHA EVELEIGH |
Manufacturer Street | ONE EDWARDS WAY |
Manufacturer City | IRVINE CA 92614 |
Manufacturer Country | US |
Manufacturer Postal | 92614 |
Manufacturer Phone | 9492503939 |
Manufacturer G1 | EDWARDS LIFESCIENCES LLC |
Manufacturer Street | ONE EDWARDS WAY |
Manufacturer City | IRVINE CA 92614 |
Manufacturer Country | US |
Manufacturer Postal Code | 92614 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VIGILEO |
Generic Name | COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION |
Product Code | DXG |
Date Received | 2020-02-13 |
Model Number | MHM1 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | EDWARDS LIFESCIENCES |
Manufacturer Address | ONE EDWARDS WAY IRVINE CA 92614 US 92614 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-13 |