NEOPUFF INFANT RESUSCITATOR RD900

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-13 for NEOPUFF INFANT RESUSCITATOR RD900 manufactured by Fisher & Paykel Healthcare Ltd.

Event Text Entries

[184420424] (b)(4). The complaint rd900 neopuff infant resuscitator was not returned to fisher & paykel healthcare (f&p) for evaluation. It is most likely that the reported broken outlet port was the result of impact. The neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infant until adequate spontaneous breathing occurs. Being a portable device, the neopuff can be susceptible to damage, for instance when accidentally dropped or subjected to considerable external force. Each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications. Any unit that fails is rejected. The subject neopuff would have met the required specification at the time of production. The neopuff technical manual warns against dropping the neopuff or subjecting it to impact damage which may cause the unit to operate incorrectly. The neopuff technical manual also states the following: dropping the neopuff infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit. If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient.
Patient Sequence No: 1, Text Type: N, H10


[184420425] A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that the gas outlet port of rd900 neopuff infant resuscitator was broken. There was no reported patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611451-2020-00154
MDR Report Key9709849
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-13
Date of Report2020-01-17
Date Mfgr Received2020-01-17
Device Manufacturer Date2010-03-24
Date Added to Maude2020-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS FARANAK GOMAROONI
Manufacturer Street173 TECHNOLOGY DRIVE SUITE 100
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9194534000
Manufacturer G1FISHER & PAYKEL HEALTHCARE LTD
Manufacturer Street15 MAURICE PAYKEL PLACE EAST TAMAKI
Manufacturer CityAUCKLAND, 2013
Manufacturer CountryNZ
Manufacturer Postal Code2013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEOPUFF INFANT RESUSCITATOR
Generic NameBTL
Product CodeBTL
Date Received2020-02-13
Model NumberRD900
Catalog NumberRD900
Lot Number100324
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFISHER & PAYKEL HEALTHCARE LTD
Manufacturer Address15 MAURICE PAYKEL PLACE EAST TAMAKI AUCKLAND, 2013 NZ 2013


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.