ZERO TIP M0063901050 390-105

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-14 for ZERO TIP M0063901050 390-105 manufactured by Boston Scientific Corporation.

Event Text Entries

[179405666] Medwatch #: mw5091277. Patient identifier: (b)(6). Patient weight: (b)(6) kg. The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown. (b)(4). According to the complainant, the suspect device has been disposed and is not available for return. If any further relevant information is received, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[179405667] It was reported to boston scientific corporation that a zero tip stone retrieval basket was used during a ureteroscopy procedure performed on (b)(6) 2019. According to the complainant, during the procedure, the basket was gently pulled on with constant pressure but the middle of the shaft broke into the upper ureter. The stone was unable to be released from the basket, and an attempt to retrieve with a grasper was unsuccessful. Two basket wires were then lasered in order to release the stone. The patient needed two additional procedures to remove the retained basket wires. Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2020-00439
MDR Report Key9710834
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-14
Date of Report2020-02-14
Date of Event2019-10-25
Date Mfgr Received2020-01-21
Date Added to Maude2020-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZERO TIP
Generic NameDISLODGER, STONE, BASKET, URETERAL, METAL
Product CodeFFL
Date Received2020-02-14
Model NumberM0063901050
Catalog Number390-105
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-14

© 2026 FDA.report