3M? RANGER? HIGH FLOW DISPOSABLE SET N/A 24355

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2020-02-14 for 3M? RANGER? HIGH FLOW DISPOSABLE SET N/A 24355 manufactured by 3m Health Care.

Event Text Entries

[179547930] No information was provided. No patient/staff injury was alleged. Initial reporter's occupation is unknown. The sample has not been received by 3m st. Paul, mn for evaluation. Analysis is pending upon receipt of product. 3m is unable to verify the leak, identify exact location and root cause without the sample. Based on the information and product photos provided, the reported incident appears to be a solvent bond leak of the y-connector. Lot hx8321 was produced post corrective action of solvent application and volume uniformity. 3m will continue to monitor. End of report.
Patient Sequence No: 1, Text Type: N, H10


[179547931] A hospital employee alleged a 3m? Ranger? High flow disposable set (model 24355) leaked saline fluid at the y-connection during priming on (b)(6) 2020 in the operating room. No pressure device or pump was used. New tubing and saline were used to prime a second set up. No patient or staff injury was alleged.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2110898-2020-00011
MDR Report Key9712575
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2020-02-14
Date of Report2020-02-14
Date of Event2020-01-09
Date Mfgr Received2020-01-30
Device Manufacturer Date2019-02-07
Date Added to Maude2020-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DIANNE GIBBS
Manufacturer Street3M CENTER, BUILDING 275-5W-06 2510 CONWAY AVE
Manufacturer CityST. PAUL, MN
Manufacturer CountryUS
Manufacturer Phone7379117
Manufacturer G1NYPRO PRECISION ASSEMBLIES
Manufacturer StreetSOR JUANA INES DE LA CRUZ CD. INDUSTRIAL
Manufacturer CityTIJUANA, 22444
Manufacturer CountryMX
Manufacturer Postal Code22444
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M? RANGER? HIGH FLOW DISPOSABLE SET
Generic NameHIGH FLOW DISPOSABLE TUBING
Product CodeBSB
Date Received2020-02-14
Model NumberN/A
Catalog Number24355
Lot NumberHX8321
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer Address2510 CONWAY AVE ST. PAUL, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-14

© 2026 FDA.report