3M? RANGER? HIGH FLOW DISPOSABLE SET N/A 24355

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2020-02-14 for 3M? RANGER? HIGH FLOW DISPOSABLE SET N/A 24355 manufactured by 3m Health Care.

Event Text Entries

[182294031] Patient information: no information was provided. No patient or staff injury was alleged. Date of event: exact date of event not provided. Reporter indicated event occurred in (b)(6) 2020. The sample was not available for evaluation. 3m is unable to verify the leak, identify exact location and root cause without the sample. Based on the provided information, the reported incident appears to have occurred due to solvent bond failure. 3m will continue to monitor.
Patient Sequence No: 1, Text Type: N, H10


[182294032] A hospital crna in (b)(6) reported that in (b)(6) 2020 the anesthesia department discovered a 3m? Ranger? High flow disposable set (model 24355) leaked ringer's solution fluid directly below the drip chamber. A 3m? Ranger? Warming unit (model 24510) was used to warm the set. No pressure device or pump was used. No patient or staff injury was alleged.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2110898-2020-00012
MDR Report Key9712577
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2020-02-14
Date of Report2020-02-14
Date of Event2020-01-01
Date Mfgr Received2020-01-24
Date Added to Maude2020-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DIANNE GIBBS
Manufacturer Street3M CENTER, BUILDING 275-5W-06 2510 CONWAY AVE
Manufacturer CityST. PAUL, MN
Manufacturer CountryUS
Manufacturer Phone7379117
Manufacturer G1NYPRO PRECISION ASSEMBLIES
Manufacturer StreetSOR JUANA INES DE LA CRUZ CD. INDUSTRIAL
Manufacturer CityTIJUANA, 22444
Manufacturer CountryMX
Manufacturer Postal Code22444
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M? RANGER? HIGH FLOW DISPOSABLE SET
Generic NameHIGH FLOW DISPOSABLE TUBING
Product CodeBSB
Date Received2020-02-14
Model NumberN/A
Catalog Number24355
Lot NumberHX8341
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer Address2510 CONWAY AVE ST. PAUL, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-14

© 2026 FDA.report