STRAIGHT SUCTION ENT 9733449

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-14 for STRAIGHT SUCTION ENT 9733449 manufactured by Medtronic Navigation, Inc.

Event Text Entries

[181947279] The device is currently lost and not available for analysis. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[181947280] Medtronic received information regarding a navigation system. It was reported that intra-operatively during a functional endoscopic sinus surgery (fess) procedure it was difficult to verify the straight suction. When the straight suction was tried to be registered (verified), it was hard to do so. Registration was not possible unless the tip was slightly shifted from the divot. It was confirmed it was bent and the reason for the suction damage was due to deterioration. Another suction was used and navigation was continued. There was no delay to the procedure and no reported impact on patient outcome.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1723170-2020-00516
MDR Report Key9714303
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-14
Date of Report2020-02-14
Date of Event2019-09-27
Date Mfgr Received2020-01-19
Date Added to Maude2020-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC
Manufacturer Street826 COAL CREEK CIRCLE
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal Code80027
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTRAIGHT SUCTION ENT
Generic NameEAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Product CodePGW
Date Received2020-02-14
Model Number9733449
Catalog Number9733449
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC
Manufacturer Address826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.