NIPRO BLOOD TUBING SET W/TRANSDUCER PROTECTOR BL+A209Y/V803

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-14 for NIPRO BLOOD TUBING SET W/TRANSDUCER PROTECTOR BL+A209Y/V803 manufactured by Nipro (thailand) Corp. Ltd..

Event Text Entries

[188104641] Noted blood leaking on a patient blood tubing set, at the blood pump area, upon inspecting, noted a tear on the blood pump segment of the blood tubing set and blood was leaking, about 250cc total blood loss. Device lot number was not recorded. Notified biomed, he noted that the blood pump rotor was missing the guides for the blood lines and may or may not have caused the tear. Biomed was able to change the malfunctioned part, the treatment was resumed with new lines, the patient was asymptomatic, vital signs were stable, and patient was able to finish treatment without problems, facility administrator and physician was made aware of the event. The patient did return for their next scheduled treatment without issues.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8041145-2019-00019
MDR Report Key9714612
Report SourceUSER FACILITY
Date Received2020-02-14
Date of Report2020-02-14
Date of Event2019-12-09
Date Mfgr Received2019-12-27
Date Added to Maude2020-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MICHELLE TEJADA
Manufacturer Street3150 NW 107 AVENUE
Manufacturer CityMIAMI, FL
Manufacturer CountryUS
Manufacturer Phone5997174234
Manufacturer G1NIPRO (THAILAND) CORP. LTD.
Manufacturer Street10/2 MOO 8, BANGNOMKO, SENA
Manufacturer CityAYUTHAYA,?, 13110
Manufacturer CountryTH
Manufacturer Postal Code13110
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNIPRO BLOOD TUBING SET W/TRANSDUCER PROTECTOR
Generic NameBLOODLINES
Product CodeFJK
Date Received2020-02-14
Returned To Mfg2020-01-21
Model NumberBL+A209Y/V803
Catalog NumberBL+A209Y/V803
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNIPRO (THAILAND) CORP. LTD.
Manufacturer Address10/2 MOO 8, BANGNOMKO, SENA AYUTHAYA,?, US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-02-14

© 2024 FDA.report