BRK? TRANSSEPTAL NEEDLE, 71 CM LENGTH 407200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-14 for BRK? TRANSSEPTAL NEEDLE, 71 CM LENGTH 407200 manufactured by St. Jude Medical.

Event Text Entries

[183003163] The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.?
Patient Sequence No: 1, Text Type: N, H10


[183003164] Related manufacturing ref: 3005334138-2020-00060. During the procedure, a dissection of the superior vena cava (svc) occurred. While preparing the introducer prior to patient insertion, a small obstruction was observed at the tip of the sheath while flushing. The brk needle was inserted through the sheath to clear the blockage but resistance was noted. Despite the obstruction and insertion difficulty, the introducer set and brk were inserted into the patient. Once positioned at the svc, the brk needle was attempted to be inserted through the introducer however resistance was felt. The needle was pushed through the introducer and dissected the svc. The introducer and needle were removed and ice confirmed a small bleed. The bleed resolved without any intervention. After the bleeding resolved the procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008452825-2020-00094
MDR Report Key9714841
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-14
Date of Report2020-03-17
Date of Event2020-01-23
Date Mfgr Received2020-03-11
Device Manufacturer Date2019-07-19
Date Added to Maude2020-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL
Manufacturer CityALAJUELA, COSTA RICA 1897-4050
Manufacturer CountryCS
Manufacturer Postal Code1897-4050
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBRK? TRANSSEPTAL NEEDLE, 71 CM LENGTH
Generic NameTROCAR
Product CodeDRC
Date Received2020-02-14
Returned To Mfg2020-02-25
Model Number407200
Catalog Number407200
Lot Number7096452
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL
Manufacturer AddressPARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL ALAJUELA, COSTA RICA 1897-4050 CS 1897-4050


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-02-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.