CHEST STAND C-9800R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2007-12-18 for CHEST STAND C-9800R manufactured by Trex Medical Inc.

Event Text Entries

[773761] Hologic received a report in 2007 from philips, who functions as a service provider for the hosp that owns the unit. The report stated that during positioning of a pt for a chest exam, the radiation technologist adjusted the pt support device, and the device released by itself and struck the pt on the head. The service engineer from philips found broken tension spring in the support device. Since this prod is no longer supported, and parts are not available to repair the equipment, the service engineer removed the pt support device from service.
Patient Sequence No: 1, Text Type: D, B5


[7929893] Hologic contacted the radiological tech. The radiological tech told us that the injury to the pt related to this incident was very minor, and no medical intervention was required. The device could not be repaired, as parts are no longer available. The pt support device was removed from service. Trex medical is no longer in business, and the device is no longer manufactured.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1221300-2007-00007
MDR Report Key971522
Report Source00
Date Received2007-12-18
Date of Report2007-12-11
Date of Event2007-10-29
Date Mfgr Received2007-11-14
Device Manufacturer Date1999-10-01
Date Added to Maude2008-04-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactARTHUR FRIEDMAN, VP
Manufacturer Street35 CROSBY DR.
Manufacturer CityBEDFORD MA 01730
Manufacturer CountryUS
Manufacturer Postal01730
Manufacturer Phone7819997368
Manufacturer G1TREX MEDICAL
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHEST STAND
Generic NameX-RAY DEVICE
Product CodeIYB
Date Received2007-12-18
Model NumberC-9800R
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key974373
ManufacturerTREX MEDICAL INC


Patients

Patient NumberTreatmentOutcomeDate
10 2007-12-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.