MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-17 for SWAN-GANZ BIPOLAR PACING CATHETER (FOR INSERTION VIA SUPERIOR VENA CAVA) D97120F5 manufactured by Edwards Lifesciences, Pr.
[188239137]
The product is expected to be returned for analysis; however, it has not yet been received. Upon the return of the product a supplemental report will be sent with the investigation results. The lot number was not provided; therefore, a review of the manufacturing records could not be completed.
Patient Sequence No: 1, Text Type: N, H10
[188239138]
It was reported that after insertion of a pacing swan-ganz catheter, the wire was noted to have migrated. The contamination shield did not lock the catheter proximally and distally to the sheath; therefore, they were unable to tighten the wire to secure it and it unintentionally migrated 3cm, which caused pacing difficulty. The contamination shield and the catheter were removed. It is unknown whether a new catheter was then used. There was no allegation of patient injury. Patient demographics were requested and are unknown. The lot number was unknown since the packaging had been discarded.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2015691-2020-10565 |
MDR Report Key | 9717211 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-02-17 |
Date of Report | 2020-02-05 |
Date of Event | 2020-02-04 |
Date Mfgr Received | 2020-02-05 |
Date Added to Maude | 2020-02-17 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. LYNN THOMAS |
Manufacturer Street | ONE EDWARDS WAY |
Manufacturer City | IRVINE CA 92614 |
Manufacturer Country | US |
Manufacturer Postal | 92614 |
Manufacturer Phone | 9497564386 |
Manufacturer G1 | EDWARDS LIFESCIENCES, PR |
Manufacturer Street | STATE RD INDUS PK 402 KM 1.4 |
Manufacturer City | ANASCO PR 00610 |
Manufacturer Country | US |
Manufacturer Postal Code | 00610 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SWAN-GANZ BIPOLAR PACING CATHETER (FOR INSERTION VIA SUPERIOR VENA CAVA) |
Generic Name | CATHETER, FLOW DIRECTED |
Product Code | DYG |
Date Received | 2020-02-17 |
Returned To Mfg | 2020-04-06 |
Model Number | D97120F5 |
Catalog Number | D97120F5 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | EDWARDS LIFESCIENCES, PR |
Manufacturer Address | STATE RD INDUS PK 402 KM 1.4 ANASCO PR 00610 US 00610 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-17 |