NIM? 3.0 INTERFACE 8253200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-18 for NIM? 3.0 INTERFACE 8253200 manufactured by Medtronic Xomed Inc..

Event Text Entries

[180612989] Analysis found that the unit arrived with channel 2 failure due to loose cable connector. Reseated the connector and replaced worn wave washer. The item was tested and passed all manufacturing specifications. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[180612990] A manufacturer representative was called on by a health are provider (hcp) due to concern on the interface. Upon checking by the rep, the device would not detect stimulus or response, the fuse was changed and repeated with the same outcome. There was no patient or staff impact.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-2020-00092
MDR Report Key9719703
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-18
Date of Report2020-02-18
Date of Event2019-12-18
Date Mfgr Received2020-01-21
Device Manufacturer Date2018-07-05
Date Added to Maude2020-02-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNIM? 3.0 INTERFACE
Generic NameSTIMULATOR, NERVE
Product CodeETN
Date Received2020-02-18
Returned To Mfg2020-02-12
Model Number8253200
Catalog Number8253200
Lot Number215769532
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.