NAVICARE? NURSE CALL SYSTEMS GRS-5

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-02-18 for NAVICARE? NURSE CALL SYSTEMS GRS-5 manufactured by Hill-rom, Inc..

Event Text Entries

[179745306] Due to instances of failed paging our communications department conducted a study. This study and one conducted by hill-rom confirmed that the grs-5 monitor spikes at the frequency 152. 0075. This is an fcc licensed frequency for medical pagin. The display gives out roughly -70db of radio frequency which is enough to interfere with paging in a 6-8ft radius based on our testing. Hill-rom was informed and did make recommendations for our testing. Spectrum analysis was done and test pages were conducted at stations with and without displays across the hospital, in a variety of environments and locations. It was found that all stations with displays in all settings interfere with the medical paging frequency. Hill-rom recommends not to use 152. 0075 for paging. However, this is a licensed frequency for medical paging and our infrastructure is built to facilitate this. It is my opinion that devices should not be introduced to a medical setting which interfere with this frequency. Manufacturer response for display, hill-rom (per site reporter). Hill rom engaged in testing of their own and have shared some of the results with us. They have released a recommendation to change our frequency for paging.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9719946
MDR Report Key9719946
Date Received2020-02-18
Date of Report2015-10-21
Date of Event2015-09-22
Report Date2020-02-04
Date Reported to FDA2020-02-04
Date Reported to Mfgr2020-02-18
Date Added to Maude2020-02-18
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNAVICARE? NURSE CALL SYSTEMS
Generic NameSYSTEM, COMMUNICATION, POWERED
Product CodeILQ
Date Received2020-02-18
Model NumberGRS-5
Device AvailabilityY
Device Age1 YR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM, INC.
Manufacturer Address1225 CRESCENT GREEN DR. SUITE 300 CARY NC 27518 US 27518


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.