CWS 400 CLOSED WOUND SUCTION KIT, PVC DRAIN, 12.5" (31.7CM) 0043610

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other,user facility report with the FDA on 2020-02-18 for CWS 400 CLOSED WOUND SUCTION KIT, PVC DRAIN, 12.5" (31.7CM) 0043610 manufactured by C.r. Bard, Inc. (covington) -1018233.

Event Text Entries

[183973712] The investigation is still in progress. Once the investigation is complete, a supplemental report will be filed. The device was not returned.
Patient Sequence No: 1, Text Type: N, H10


[183973713] It was reported that the patient's drain was noted to be disconnected at the "lobster claw" area once he sat on the edge of his bed. "neurosurgery provider advised to wipe tubing clean with alcohol swab, reconnect, and recompress drain. Did not recommend removing drain completely as patients surgery was just the day before. " the patient also got out of bed to sit in his chair a few hours later. Drain was clamped to his gown with the green clasp. The clasp let loose and the drain fell to the floor. The tubing became disconnected from the actual drain. It was wiped with alcohol and reconnected. The drain was reportedly in use for 13 hours.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1018233-2020-01085
MDR Report Key9721250
Report SourceOTHER,USER FACILITY
Date Received2020-02-18
Date of Report2020-03-10
Date of Event2020-01-11
Date Mfgr Received2020-03-04
Date Added to Maude2020-02-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYONIC ANDERSON
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal30014
Manufacturer Phone7707846100
Manufacturer G1C.R. BARD, INC. (COVINGTON) -1018233
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal Code30014
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCWS 400 CLOSED WOUND SUCTION KIT, PVC DRAIN, 12.5" (31.7CM)
Generic NameWOUND DRAIN
Product CodeGCY
Date Received2020-02-18
Catalog Number0043610
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerC.R. BARD, INC. (COVINGTON) -1018233
Manufacturer Address8195 INDUSTRIAL BLVD COVINGTON GA 30014 US 30014


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.