EMBOSHIELD NAV 6 22437-19

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-02-18 for EMBOSHIELD NAV 6 22437-19 manufactured by Abbott Vascular.

Event Text Entries

[180155486] The device was received. Investigation is not yet complete. A follow up report will be submitted with all additional relevant information.
Patient Sequence No: 1, Text Type: N, H10


[180155487] It was reported that prior to use and during prep, the shaft of the emboshield nav6 embolic protection system (eps) kinked when the filter was being loaded into the delivery catheter (dc) pod. No resistance was noted and there was no bunching or wrinkling noted on the dc pod. The device was not used and there was no patient involvement. Another emboshield nav6 was used to complete the procedure. There was no clinically significant delay in the procedure. Return device analysis identified a separation of the distal tip of the retrieval catheter. The account confirmed that the separation occurred during loading of the filter. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-01578
MDR Report Key9721800
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-02-18
Date of Report2020-03-24
Date of Event2019-11-25
Date Mfgr Received2020-02-27
Device Manufacturer Date2019-06-17
Date Added to Maude2020-02-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 2024168
Manufacturer Street26531 YNEZ ROAD
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal Code925914628
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEMBOSHIELD NAV 6
Generic NameEMBOLIC PROTECTION SYSTEM
Product CodeNTE
Date Received2020-02-18
Returned To Mfg2020-01-02
Model Number22437-19
Catalog Number22437-19
Lot Number9061761
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.