[179920879] At time of filing, although expected, the reported device has not been returned to conmed for evaluation. A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation. This issue will continue to be monitored through the complaint system to assure patient safety.
Patient Sequence No: 1, Text Type: N, H10


[179920880] On behalf of the customer, the conmed sales representative submitted photos and reported an issue involving the l3000, lithium battery charger sn (b)(4) that occurred at (b)(6) hospital. It was noted when charging a battery l3000lg on (b)(6) 2020 the charger caught on fire. There were no outside sources that could have caused the fire. Information from the facility indicates that no one was injured, there was no patient or surgical involvement. The charger and battery were located within the central sterile processing room. Only one of the four connection points was involved with the fire. The battery had not been on the charger any longer than 2-3 hours as they rotate the batteries frequently during use. It was stated the charger is kept in use throughout the day, and as they process batteries through the cycles they will rotate the batteries onto the charger. As per witnesses, the flames originated from inside of the charger, and stemmed upwards through the battery which then caused the melting of the battery and surrounding area on the charger itself. There were flames also coming from the charging station slot #1 where the battery was connected. The staff utilized a fire extinguisher to put out the flames. The other charging stations had been in use when the accident occurred but there was no evident or visible damage to the other batteries on the charging station. Those other batteries were separated from the rotation and were sent out with the other pieces of equipment as precaution. There were no visible damages to the cord before or after the fire, but it was sent out for evaluation as well. This report is being raised on the basis of malfunction with potential for injury upon reoccurrence due to fire.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017294-2020-00072
MDR Report Key9722659
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-18
Date of Report2020-02-28
Date of Event2020-01-30
Date Mfgr Received2020-02-12
Device Manufacturer Date2014-01-28
Date Added to Maude2020-02-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS TRACEY WEISELBENTON
Manufacturer Street11311 CONCEPT BLVD
Manufacturer CityLARGO, FL
Manufacturer CountryUS
Manufacturer Phone3995557
Manufacturer G1CONSOLIDATED MEDICAL EQUIPMENT
Manufacturer StreetALEJANDRO DUMAS AVE, 11321 COMPLEJO INDUSTRIAL CHIHUAHUA
Manufacturer CityCHIHUAHUA, MEXICO 31136
Manufacturer CountryMX
Manufacturer Postal Code31136
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHALL LITHIUM BATTERY CHARGER
Generic NameBATTERY, REPLACEMENT, RECHARGEABLE
Product CodeMOQ
Date Received2020-02-18
Returned To Mfg2020-02-12
Catalog NumberL3000
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCONSOLIDATED MEDICAL EQUIPMENT
Manufacturer AddressALEJANDRO DUMAS AVE, 11321 COMPLEJO INDUSTRIAL CHIHUAHUA CHIHUAHUA, MEXICO 31136 MX 31136


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.