[180723042] Vyaire technician cleaned the ventilator, replaced driver, installed preventive maintenance kit and calibrated the unit. Ventilator passed all performance check and no problem found.
Patient Sequence No: 1, Text Type: N, H10


[180723043] The customer reported that their 3100a ventilator was involved in an incident while ventilating on a patient, they removed the ventilator and suctioned the patient. When they reconnected the ventilator to the patient, the unit would not start. The ventilator pressurized and started when the stopper was in circuit; they reconnect it to the patient and the ventilator did not start. No patient harm was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2020-11372
MDR Report Key9722850
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-02-18
Date of Report2020-01-21
Date of Event2020-01-21
Date Mfgr Received2020-01-21
Device Manufacturer Date1999-06-28
Date Added to Maude2020-02-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMINDY FABER
Manufacturer Street26125 N. RIVERWOODS BLVD.
Manufacturer CityMETTAWA, IL
Manufacturer CountryUS
Manufacturer Phone7570116
Manufacturer G1VYAIRE MEDICAL INC.
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV
Generic NameVENTILATOR, HIGH FREQUENCY
Product CodeLSZ
Date Received2020-02-18
Model Number3100A
Catalog Number768901
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL
Manufacturer Address26125 N. RIVERWOODS BLVD. METTAWA, IL US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.