3M? RED DOT? MONITORING ELECTRODE WITH FOAM TAPE AND STICKY GEL N/A 2570

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-02-19 for 3M? RED DOT? MONITORING ELECTRODE WITH FOAM TAPE AND STICKY GEL N/A 2570 manufactured by 3m Health Care.

Event Text Entries

[182295400] Patient information: no information was provided. Date of event: exact date of event was not provided. Date received by manufacturer was used as the d. O. E. The sample was not returned for evaluation. 3m will continue to monitor.
Patient Sequence No: 1, Text Type: N, H10


[182295401] A hospital employee reported that 3m? Red dot? Monitoring electrodes with foam tape and sticky gel, model 2570, are getting stuck in the lead wires, where the electrode buttons came off the electrode pad and stayed lodged in the lead snap. A non-3m brand of ecg lead wires was used by the facility. No patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2110898-2020-00007
MDR Report Key9727401
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-02-19
Date of Report2020-02-19
Date of Event2020-01-22
Date Mfgr Received2020-01-22
Date Added to Maude2020-02-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DIANNE GIBBS
Manufacturer Street3M CENTER, BUILDING 275-5W-06 2510 CONWAY AVE
Manufacturer CityST. PAUL, MN
Manufacturer CountryUS
Manufacturer Phone7379117
Manufacturer G13M CANADA COMPANY
Manufacturer Street400 ROUTE 100
Manufacturer CityMORDEN, R6M 1Z9
Manufacturer CountryCA
Manufacturer Postal CodeR6M 1Z9
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M? RED DOT? MONITORING ELECTRODE WITH FOAM TAPE AND STICKY GEL
Generic NameECG ELECTRODE
Product CodeDRX
Date Received2020-02-19
Model NumberN/A
Catalog Number2570
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer Address2510 CONWAY AVE ST. PAUL, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.