POWERPRO STERILE TRANSFER BATTERY CASE PRO3110

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,07 report with the FDA on 2007-12-18 for POWERPRO STERILE TRANSFER BATTERY CASE PRO3110 manufactured by Conmed Linvatec.

Event Text Entries

[747823] The customer reported that during use of this battery case in a total arthroplasty surgery, the rails broke. The user reported notable vibration of the handpiece in use. This event resulted in the battery case falling off of the handpiece and pieces of the rail falling into the surgical site. The customer reported difficulty removing the rail fragments from the surgical site but this did not result in injury to the patient.
Patient Sequence No: 1, Text Type: D, B5


[7902825] Investigation results: conmed linvatec is unable to confirm this reported problem because the device is not being returned for evaluation. The instructions for use contain the following warning: inspect the sterile transfer battery case for any damage, such as: cracks in the case, or damaged, broken, or worn o-rings, latch and/or hinge. If damage is found, do not use. Sterility may be compromised and patient infection may occur. A corrective action is in process to address this failure mode. Conmed linvatec will continue to monitor this device for failures.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017294-2007-00848
MDR Report Key972777
Report Source01,05,07
Date Received2007-12-18
Date of Report2007-11-23
Date of Event2007-11-20
Date Mfgr Received2007-11-23
Date Added to Maude2008-04-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBEVERLY SCHANER
Manufacturer Street11311 CONCEPT BLVD.
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal33773
Manufacturer Phone7273995169
Manufacturer G1CONMED LINVATEC
Manufacturer Street11311 CONCEPT BLVD.
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal Code33773
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePOWERPRO STERILE TRANSFER BATTERY CASE
Generic NameBATTERY CASE
Product CodeFCO
Date Received2007-12-18
Catalog NumberPRO3110
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key975424
ManufacturerCONMED LINVATEC
Manufacturer AddressLARGO FL US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-12-18