CADD LEGACY PCA 6300 21-6300-51

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-19 for CADD LEGACY PCA 6300 21-6300-51 manufactured by Smiths Medical Asd, Inc.

Event Text Entries

[180112198] Evaluation results: one cadd legacy pca was returned for investigation in used condition. The housing was damaged and the screen was scratched. Error code 1310 was observed upon power up. The error was cleared. The problem source of the reported product problem was unknown. A root cause was not established.
Patient Sequence No: 1, Text Type: N, H10


[180112199] It was reported that the pump stopped infusing due to an alarm. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-00023
MDR Report Key9728780
Report SourceUSER FACILITY
Date Received2020-02-19
Date of Report2020-02-19
Date Mfgr Received2019-12-19
Device Manufacturer Date2008-03-07
Date Added to Maude2020-02-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer G1SMITHS MEDICAL ASD, INC
Manufacturer Street3350 GRANADA AVENUE NORTH SUITE 100
Manufacturer CityOAKDALE, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCADD LEGACY PCA
Generic NamePUMP, INFUSION, PCA
Product CodeMEA
Date Received2020-02-19
Returned To Mfg2019-12-30
Model Number6300
Catalog Number21-6300-51
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-19

© 2026 FDA.report