RAPIDLAB 1265 BLOOD GAS ANALYZER 10321852

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-02-19 for RAPIDLAB 1265 BLOOD GAS ANALYZER 10321852 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[180902972] Siemens reviewed the data logs provided by the customer. The co-oximetry functionality on the rl1265 appears to have been working as intended throughout the day of and including when the sample in question was measured. This is concluded based on the co-ox calibration stability, internal co-ox quality checks, aqc expected recovery and lack of any relevant diagnostic errors or exception flags. The reported thb values appear valid. It is known that for accurate thb measurements, whole blood specimens require the red blood cells to be uniformly distributed throughout the entire sample. This can be obtained only by thorough mixing of the blood sample across two planes performed immediately before each analysis. Service discussed the importance of pre-analytical concerns with the customer. The instrument is operational. The cause of this event is unknown.
Patient Sequence No: 1, Text Type: N, H10


[180902973] The customer reported discrepant total hemoglobin results on the rl 1265. There was no report of injury due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002637618-2020-00004
MDR Report Key9729533
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-02-19
Date of Report2020-02-19
Date of Event2020-01-30
Date Mfgr Received2020-02-03
Date Added to Maude2020-02-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactFELIX AKINRINOLA
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD, MA
Manufacturer CountryUS
Manufacturer Phone7052212
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
Manufacturer StreetNORTHERN ROAD CHILTON INDUSTRIAL ESTATE
Manufacturer CitySUDBURY, CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRAPIDLAB 1265 BLOOD GAS ANALYZER
Generic NameRL 1265
Product CodeCHL
Date Received2020-02-19
Model Number10321852
Catalog Number10321852
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN, NY US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-19

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.