TEG? HEMOSTASIS ANALYZER SYSTEM TEG 6S UNIT PACK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-02-20 for TEG? HEMOSTASIS ANALYZER SYSTEM TEG 6S UNIT PACK manufactured by Haemonetics Corporation.

Event Text Entries

[180167282] Results were achieved after 4 cartridges were used. Customer indicated to haemonetics that delay in processing was not responsible for patient death. A field service engineer was dispatched and inspected the teg6s and checked the calibration settings of the unit and found no issues. Ran qc levels 1 and 2. Unit meets manufacturer specifications.
Patient Sequence No: 1, Text Type: N, H10


[180167283] On (b)(6) 2020, customer informed haemonetics of a patient fatality that happened at (b)(6) medical center in (b)(6). Palette mapping, on a teg6s, gave a sudlite error in the middle of testing and there was a delay in obtaining the results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219343-2020-00012
MDR Report Key9733033
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-02-20
Date of Report2020-01-22
Date of Event2019-12-27
Date Mfgr Received2020-01-22
Device Manufacturer Date2016-08-10
Date Added to Maude2020-02-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR SHAUN FLANAGAN
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON, MA
Manufacturer CountryUS
Manufacturer G1HAEMONETICS CORPORATION
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON, MA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTEG? HEMOSTASIS ANALYZER SYSTEM
Generic NameTEG 6S US UNIT PACK
Product CodeJOZ
Date Received2020-02-20
Model NumberTEG 6S UNIT PACK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHAEMONETICS CORPORATION
Manufacturer Address125 SUMMER STREET BOSTON, MA US


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2020-02-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.