BARD? CLOSED SYSTEM URINARY DRAINAGE BAG 153509

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,other report with the FDA on 2020-02-20 for BARD? CLOSED SYSTEM URINARY DRAINAGE BAG 153509 manufactured by C.r. Bard, Inc. (covington) -1018233.

Event Text Entries

[180332105] The reported event was inconclusive, as the device was not returned for evaluation. However, a potential failure mode could be? Bag will not expand/vinyl stuck together? With a potential root cause of? Static or positive air pressure.? The lot number is unknown; therefore, the device history record could not be reviewed. The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use. If package is opened or if any imperfection or surface deterioration is observed, do not use. "
Patient Sequence No: 1, Text Type: N, H10


[180332106] It was reported that when the patient moved the bag around, the urine came back up into the tube. The complainant confirmed that the bag was not being held upside down.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1018233-2020-01225
MDR Report Key9734638
Report SourceDISTRIBUTOR,OTHER
Date Received2020-02-20
Date of Report2020-02-20
Date Mfgr Received2020-01-30
Date Added to Maude2020-02-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYONIC ANDERSON
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal30014
Manufacturer Phone7707846100
Manufacturer G1C.R. BARD, INC. (COVINGTON) -1018233
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal Code30014
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBARD? CLOSED SYSTEM URINARY DRAINAGE BAG
Generic NameDRAIN BAG
Product CodeKNX
Date Received2020-02-20
Model Number153509
Catalog Number153509
Lot NumberUNK
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerC.R. BARD, INC. (COVINGTON) -1018233
Manufacturer Address8195 INDUSTRIAL BLVD COVINGTON GA 30014 US 30014


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.