NCIRCLE TIPLESS STONE EXTRACTOR NTSE-022115-UDH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-21 for NCIRCLE TIPLESS STONE EXTRACTOR NTSE-022115-UDH manufactured by Cook Inc.

Event Text Entries

[183797626] Initial reporter: postal code: (b)(6). Pma/510k # exempt. This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[183797627] It was reported that during a flexible ureteroscopy of the urinary tract, a ncircle tipless stone extractor did not function "properly". This occurred prior to direct patient contact. Another same device was used to complete the procedure. It was reported that the patient did not experience any adverse effects due to this alleged malfunction.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2020-00414
MDR Report Key9738986
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-21
Date of Report2020-03-17
Date of Event2020-02-03
Date Mfgr Received2020-03-16
Device Manufacturer Date2019-11-01
Date Added to Maude2020-02-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. JENNIFER CANADA
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNCIRCLE TIPLESS STONE EXTRACTOR
Generic NameFFL DISLODGER, STONE, BASKET, URETERAL, METAL
Product CodeFFL
Date Received2020-02-21
Returned To Mfg2020-03-11
Catalog NumberNTSE-022115-UDH
Lot Number10124274
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.