Event Text Entries

[180608970] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[180608971] A healthcare provider reported via manufacturer representative that the bur head broke off inside the patient. The hcp used a dcr bur at 30,000 rpm instead of the indicated 12,000 rpm. The bur head was removed, an x-ray was performed to check that no metal remained in the patient. The procedure was delayed for 60 minutes. The bur was stuck inside the handpiece. A different handpiece was used to finish the procedure. The patient was not injured.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045254-2020-00100
MDR Report Key9738993
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-21
Date of Report2020-02-21
Date of Event2020-01-30
Date Mfgr Received2020-01-30
Device Manufacturer Date2016-12-08
Date Added to Maude2020-02-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DR N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXPS? BLADE
Generic NameBUR, EAR, NOSE AND THROAT
Product CodeEQJ
Date Received2020-02-21
Model Number1882569HS
Catalog Number1882569HS
Lot Number0212447183
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-21