CLOSUREFAST MICRO INTRODUCER SHEATH SET 7FR/CH (2.3MM) X 7CM MIS-7F07

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-02-21 for CLOSUREFAST MICRO INTRODUCER SHEATH SET 7FR/CH (2.3MM) X 7CM MIS-7F07 manufactured by Argon Medical Devices Inc..

Event Text Entries

[184446657] The root cause has been identified as non-optimized molding parameters (melt temperatures) and work instructions that were not specific enough regarding loading of the tubing on the mandrel. A capa (reference capa (b)(4)) has been initiated to address the manufacturing error.
Patient Sequence No: 1, Text Type: N, H10


[184446658] Physician was attempting to use a micro introducer along with a rfg3 during procedure to treat the great saphenous vein (gsv). The lumen was flushed prior to use. Ifu was followed. Proper temperature was reached. A guide wire was used for the insertion of the catheter. It was reported that the blue dilator broke off the hub of the sheath, it was a clean break off the combined blue and orange hub. Since the break was clean, physician pulled blue introducer out that broke off the hub and re-opened another micro kit to complete the procedure. There was no patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1625425-2020-00174
MDR Report Key9739969
Report SourceUSER FACILITY
Date Received2020-02-21
Date of Report2020-02-03
Date of Event2020-01-29
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2020-02-03
Date Added to Maude2020-02-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. GAIL SMITH
Manufacturer Street1445 FLAT CREEK ROAD
Manufacturer CityATHENS TX 75751
Manufacturer CountryUS
Manufacturer Postal75751
Manufacturer Phone2144368995
Manufacturer G1ARGON MEDICAL DEVICES INC.
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameCLOSUREFAST MICRO INTRODUCER SHEATH SET 7FR/CH (2.3MM) X 7CM
Generic NameMICRO INTRODUCER SHEATH
Product CodeDRE
Date Received2020-02-21
Catalog NumberMIS-7F07
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerARGON MEDICAL DEVICES INC.
Manufacturer Address1445 FLAT CREEK ROAD ATHENS TX 75751 US 75751

Device Sequence Number: 101

Brand NameACETAMINOPHEN 500MG
Product Code---
Date Received2020-02-21
Device Sequence No101
Device Event Key0
ManufacturerMANUFACTURE / COMPOUNDER NAME


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-21