3M? CAVILON? NO STING BARRIER FILM 3343E N/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-21 for 3M? CAVILON? NO STING BARRIER FILM 3343E N/A manufactured by 3m Health Care Limited.

Event Text Entries

[180395207] Patient information: no information was provided. The flammability event is not due to a product defect or problem with this specific lot. The product is flammable until it is completely dry and vapors have dissipated. A flammability event will occur if the product is exposed to an ignition source before the vapors are dissipated. Based on information provided in the incident report, the ignition source for this flammability event is believed to be potential static electricity discharge. The product insert has instructions not to use cavilon no sting barrier film near sources of static discharge.
Patient Sequence No: 1, Text Type: N, H10


[180395208] A eu regulatory authority reported on (b)(6) 2019 a employee was using 3m? Cavilon? No sting barrier film 3343e in a hospital. While using the cavilon stick, the empty package was on the shelf at bedside. The employee was wearing gloves and removed the slide sheet from under a patient. The employee proceeded to pick up the empty cavilon packaging and heard a sizzle and small flames appeared. The flames went out immediately. No one was injured.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2110898-2020-00010
MDR Report Key9740776
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-21
Date of Report2020-02-21
Date of Event2019-10-01
Date Mfgr Received2019-12-06
Device Manufacturer Date2019-06-27
Date Added to Maude2020-02-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DIANNE GIBBS
Manufacturer Street3M CENTER, BUILDING 275-5W-06 2510 CONWAY AVE
Manufacturer CityST. PAUL, MN
Manufacturer CountryUS
Manufacturer Phone7379117
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M? CAVILON? NO STING BARRIER FILM 3343E
Generic Name3M? CAVILON? NO STING BARRIER FILM 3343E
Product CodeKMF
Date Received2020-02-21
Model NumberN/A
Catalog Number3343E
Lot Number2022-07 AA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE LIMITED
Manufacturer AddressDERBY ROAD LOUGHBOROUGH LEICESTERSHIRE, GB US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-21

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.