BLACKMAX-NEURO

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-02-23 for BLACKMAX-NEURO manufactured by Depuy Synthes Products Llc.

Event Text Entries

[183462670] (b)(4). This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. During repair, it was determined that the device had low power, high noise, broken vanes, and failed pretest for rotational speed and noise assessment. Therefore, the reported condition was confirmed. The assignable root cause was determined to be due to maintenance deficiency. If additional information should become available, a supplemental medwatch will be submitted accordingly.
Patient Sequence No: 1, Text Type: N, H10


[183462671] It was reported that during service and repair pretesting, it was observed that the motor device had low power. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of this event was unknown but was noted to have occurred in 2020. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1045834-2020-00312
MDR Report Key9742926
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-02-23
Date of Report2020-02-06
Date of Event2020-01-01
Date Mfgr Received2020-02-06
Device Manufacturer Date2003-08-26
Date Added to Maude2020-02-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KARA DITTY-BOVARD
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal33410
Manufacturer Phone6103142063
Manufacturer G1DEPUY SYNTHES PRODUCTS LLC
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal Code33410
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBLACKMAX-NEURO
Generic NameMOTOR, DRILL, PNEUMATIC - HANDPIECE
Product CodeHBB
Date Received2020-02-23
Returned To Mfg2016-10-21
Catalog NumberBLACKMAX-N
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES PRODUCTS LLC
Manufacturer Address4500 RIVERSIDE DRIVE PALM BEACH GARDENS FL 33410 US 33410


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-23