MICK 200, TP APPL, 20CM LONG, INSTRUMENT 8810-I-200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-02-21 for MICK 200, TP APPL, 20CM LONG, INSTRUMENT 8810-I-200 manufactured by Eckert & Ziegler Bebig / Mick Radio-nuclear Instrument, Inc..

Event Text Entries

[180851008] Mick 200 applicator malfunction x4 - serial # (b)(4) - each mick experienced a mechanical malfunction - two clip failures (determined to be from not lubricating pivot mechanism) and two seed jamming's - may be related to not performing sterile water lubrication during procedure. Less than the intended number of seeds were placed as a result. Reviewed seed placement and determined that adequate placement and dosing achieved. No further surgical intervention indicated. Pt will be followed with radiological studies and with psa labs as previously planned. Fda safety report id# (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5093218
MDR Report Key9744852
Date Received2020-02-21
Date of Report2020-02-19
Date of Event2019-12-11
Date Added to Maude2020-02-24
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameMICK 200, TP APPL, 20CM LONG, INSTRUMENT
Generic NameSYSTEM, APPLICATOR, RADIONUCLIDE, MANUAL
Product CodeIWJ
Date Received2020-02-21
Model Number8810-I-200
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerECKERT & ZIEGLER BEBIG / MICK RADIO-NUCLEAR INSTRUMENT, INC.
Manufacturer AddressMT. VERNON NY 10550 US 10550

Device Sequence Number: 2

Brand NameMICK 200, TP APPL, 20CM LONG, INSTRUMENT
Generic NameSYSTEM, APPLICATOR, RADIONUCLIDE, MANUAL
Product CodeIWJ
Date Received2020-02-21
Model Number8810-I-200
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No2
Device Event Key0
ManufacturerECKERT & ZIEGLER BEBIG / MICK RADIO-NUCLEAR INSTRUMENT, INC.

Device Sequence Number: 3

Brand NameMICK 200, TP APPL, 20CM LONG, INSTRUMENT
Generic NameSYSTEM, APPLICATOR, RADIONUCLIDE, MANUAL
Product CodeIWJ
Date Received2020-02-21
Model Number8810-I-200
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No3
Device Event Key0
ManufacturerECKERT & ZIEGLER BEBIG / MICK RADIO-NUCLEAR INSTRUMENT, INC.

Device Sequence Number: 4

Brand NameMICK 200, TP APPL, 20CM LONG, INSTRUMENT
Generic NameSYSTEM, APPLICATOR, RADIONUCLIDE, MANUAL
Product CodeIWJ
Date Received2020-02-21
Model Number8810-I-200
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No4
Device Event Key0
ManufacturerECKERT & ZIEGLER BEBIG / MICK RADIO-NUCLEAR INSTRUMENT, INC.


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-21

© 2026 FDA.report