WATER TRAP RAINTRAP W/CLP N/S 5276 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2007-12-21 for WATER TRAP RAINTRAP W/CLP N/S 5276 * manufactured by Cardinal Health.

Event Text Entries

[7930441] According to info provided by the customer, the watertrap was used with a non cardinal hlth ventilator circuit (non-heated wire) and it was noted that the clamp was in the open position at the time of the reported incident. However, there is no info of the position of the water tap in the circuit, as to whether or not it was in the upright position during use, because if the clamp were in the open position and the water trap in the upright position, the water would have drained out of the water trap and would not have accumulated in the circuit. Samples of the clamp from current inventory were reviewed in accordance with our raw material spec and no problems were found related to the clamp. Our mfg process also reviewed, and no problems were found related to the issue reported. Therefore, without the sample or add'l info we are unable to determine if the water trap contributed to the reported incident.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1423507-2007-00137
MDR Report Key974626
Report Source06
Date Received2007-12-21
Date Mfgr Received2007-11-27
Date Added to Maude2008-01-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag0
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactMICHELE DONALTCH
Manufacturer Street1430 WAUKEGAN RD. BLDG KB
Manufacturer CityMCGRAW PARK IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8475786412
Manufacturer G1CARDINAL HELATH MEXICALI
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWATER TRAP RAINTRAP W/CLP N/S
Generic NameWATER TRAP RAINTRAP W/CLP N/S
Product CodeBYH
Date Received2007-12-21
Model Number5276
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key942838
ManufacturerCARDINAL HEALTH
Manufacturer Address1500 WAUKEGAN ROAD MCGAW PARK IL 60085 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-12-21