MED/LGE CLIP APPLIER 28CM PV328-1BR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-24 for MED/LGE CLIP APPLIER 28CM PV328-1BR manufactured by Aesculap Inc.

Event Text Entries

[181347878] Manufacturing site evaluation: additional information / investigation results will be provided in a supplemental report. Investigation on-going, if available.
Patient Sequence No: 1, Text Type: N, H10


[181347879] It was reported that there was an issue with the clip appliers. During closure of an anterior lumbar interbody fusion (alif) surgery, the staff noticed that some of the color coding on one of the appliers was missing. This was part of the plastic-like overlay on the finger loop of the devices. There was flaking noted on all 3 devices and the surgeon's were concerned that this could have fallen inside the patient. No fragments were recovered and a decision was made not to reopen the incision. There was no known patient harm and no additional intervention. Associated medwatches: 2916714-2020-00044, 2916714-2020-00045, 2916714-2020-00046.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2916714-2020-00045
MDR Report Key9746862
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-24
Date of Report2020-02-24
Date of Event2020-01-28
Date Mfgr Received2020-01-28
Date Added to Maude2020-02-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSAY CHROMIAK
Manufacturer Street3773 CORPORATE PARKWAY
Manufacturer CityCENTER VALLEY, PA
Manufacturer CountryUS
Manufacturer G1AESCULAP INC
Manufacturer Street3773 CORPORATE PARKWAY
Manufacturer CityCENTER VALLEY, PA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMED/LGE CLIP APPLIER 28CM
Generic NameLAPAROSCOPIC SURGERY
Product CodeGDO
Date Received2020-02-24
Model NumberPV328-1BR
Catalog NumberPV328-1BR
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP INC
Manufacturer Address3773 CORPORATE PARKWAY CENTER VALLEY, PA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.