BIB? SYSTEM INTRAGASTRIC BALLOON B-40800

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-24 for BIB? SYSTEM INTRAGASTRIC BALLOON B-40800 manufactured by Apollo Endosurgery, Inc..

Event Text Entries

[182191435] Medwatch sent to the fda. The device was returned to the apollo device analysis laboratory on 10/feb/2020. Analysis of the device is ongoing. A review of the device labeling notes the following: the current bib? System intragastric balloon directions for use (dfu) addresses the known and anticipated potential events of "difficulty with fill tube" and "esophageal perforation" as follows: the physiological response of the patient to the presence of bib? May vary depending upon the patient's general condition and the level and type of activity. The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response. Each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications. Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms. Patients need to be evaluated and the device removed at or within 6 months of placement. Clinical data does not exist to support use of an individual bib? Beyond 6 months. Adverse events: it is important to discuss all possible complications and adverse events with your patient. Complications that may result from the use of this product include the risks associated with the medications and methods utilized in the endoscopic procedure, the risks associated with any endoscopic procedure, the risks associated with the bib? Intragastric balloon specifically, and the risks associated with the patient's degree of intolerance to a foreign object placed in the stomach. Possible complications - possible complications of the use of bib? Include: gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon. Possible complications of routine endoscopy & sedation - potential risks associated with upper endoscopic procedures include, but are not limited to: abdominal cramping and discomfort from the air used to distend the stomach, sore or irritated throat, bleeding, infection, tearing of the esophagus or stomach, and aspiration pneumonia. The risk increases if additional procedures are performed. Warnings: proper positioning of the placement catheter assembly and the bib? Balloon within the stomach (using measured distance from the incisors via the insertion tube markings) is necessary to allow for proper inflation. Lodging of the balloon in the esophageal opening during inflation may cause injury and/or device rupture. Failure to confirm proper positioning may cause injury to the esophagus, duodenum, or pylorus. Precautions: if difficulty with the bib? Fill tube is noted during placement (e. G. , resistance to balloon filling), then the device should be removed and replaced with a new balloon. To lessen, or prevent fill tube defects, during the filling process the fill tube must remain slack. If the fill tube is under tension during this process, the fill tube may dislodge from the balloon, preventing further balloon deployment. Warnings: when filling the balloon during the placement procedure, avoid rapid fill rates as these will generate high pressure which can damage the bib? Valve or cause premature detachment of the balloon from the placement catheter. Bib? Is composed of a soft silicone elastomer and is easily damaged by instruments or sharp objects. The balloon must be handled only with gloved hands and with the instruments recommended for removal. Caution: if the balloon becomes separated from the sheath prior to placement, do not attempt to use the balloon or reinsert the balloon into the sheath. Caution: during the filling process the fill tube must remain slack. If the fill tube is under tension during this process the fill tube may dislodge from the balloon preventing further balloon deployment. Warning: rapid fill rates will generate high pressure which can damage the bib? Valve or cause premature detachment.
Patient Sequence No: 1, Text Type: N, H10


[182191436] Reported as: disconnected from filling tube. Patient presented distal esophagus tear.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006722112-2020-00022
MDR Report Key9747065
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-24
Date of Report2020-01-30
Date of Event2020-01-28
Date Mfgr Received2020-02-10
Device Manufacturer Date2019-06-27
Date Added to Maude2020-02-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. DAVID HOOPER
Manufacturer Street1120 S. CAPITAL OF TEXAS HWY BLDG 1, STE. 300
Manufacturer CityCS
Manufacturer CountryUS
Manufacturer G1APOLLO ENDOSURGERY COSTA RICA, SRL
Manufacturer StreetCOYOL FREE ZONE BUILDING B 13.3
Manufacturer CityALAJUELA, CS
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBIB? SYSTEM INTRAGASTRIC BALLOON
Generic NameINTRAGASTRIC BALLOON
Product CodeLTI
Date Received2020-02-24
Returned To Mfg2020-02-10
Model NumberB-40800
Catalog NumberB-40800
Lot NumberAF03017
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAPOLLO ENDOSURGERY, INC.
Manufacturer Address1120 S. CAPITAL OF TEXAS HWY BLDG 1, STE. 300 AUSTIN, TX US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-02-24