ZOLL IVTM ICY CATHETER 8700-0782-14 IC-3893CO

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-24 for ZOLL IVTM ICY CATHETER 8700-0782-14 IC-3893CO manufactured by Zoll Circulation.

Event Text Entries

[184387901] The icy catheter was returned to zoll on (b)(6) 2020 for investigation; however, investigation is still in progress. A supplemental report will be filed when investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[184387902] After four days, during maintenance phase of 36? C on a patient using ivtm therapy, the user noticed the saline bag was low and air in the tubing. User also observed that the floor was wet. Following this, the saline bag was replaced and re-primed the start-up kit and continued with therapy. After 3 hours during continuation of the maintenance phase of therapy, the user observed an empty saline bag. The user also observed that the condensation tray was full of saline and in the air trap of the ivtm console. Ivtm therapy was discontinued. No consequences or impact to patient. Mfr 3010617000-2020-00196 for thermogard console.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010617000-2020-00195
MDR Report Key9747574
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-24
Date of Report2020-03-09
Date of Event2020-01-30
Date Mfgr Received2020-02-26
Device Manufacturer Date2019-01-14
Date Added to Maude2020-02-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KIM THOA NGUYEN
Manufacturer Street2000 RINGWOOD AVE.
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Manufacturer Phone4192922
Manufacturer G1ZOLL CIRCULATION
Manufacturer Street2000 RINGWOOD AVE.
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZOLL IVTM ICY CATHETER
Generic NameCENTRAL VENOUS CATHETER
Product CodeNCX
Date Received2020-02-24
Returned To Mfg2020-02-11
Model Number8700-0782-14
Catalog NumberIC-3893CO
Lot Number90398
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZOLL CIRCULATION